Supply Voltage:24 VAC -20%/+15% @ 5 VA, 50/60 Hz or 24 VDC -20%/+15% @ 208 mA
Signal Output:0/4-20 mA @ 500Ω; 0/2-10 VDC @ 50kΩ
Repeatability:
CO:±3 ppm of reading
NH3:3% of reading
NO2:2% of reading
O2:0.1% of reading
3000 models:±2% of reading
Measurement Range:
CO:0-250 ppm, span field calibratable 0 to 200-300
NH3:0-300 ppm
N02:0-10 ppm
O2:0-25 %
3000 models:0 to 100% L.E.L.
Sensing Technology:
1000 models:Electrochemical
3000 models:Catalytic bead
Life Expectancy:3-5 years for 1000 models, 3 years for 3000 models
Visual Indication:On, stage status, failure
Display:2 lines, 16 characters per line
Response Time:t90 <50 seconds
Relay Output:(1) SPDT. (1) NC-SPST or (1) NO SPST. 30 VAC/VDC @ 0.5 A, (10 ppm DIF) (10 ppm DIF)
Alarm Auditory Levels [Db]:83 db @ unit
Wiring Terminations:Terminal blocks, screw type for lead wire
Operating Temperature:14° to 122°F (-10° to 50°C)
Operating Humidity:15 to 95%, non-condensing
Dimensions:5.12"W x 5.12"H x 2.95“D (13 x 13 x 7.5 cm)
Weight:0.6 lbs (0.3 Kg)
Approvals:CE, City of Los Angeles
Warranty:1 year